User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 5340A Sub1 DATE: 21-Apr-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 15 NUMBER OF LINES: 85 CONFIGURATION: HP 3325A STANDARD: Mieco 1402 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 HEAD {} 1.004 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.005 HEAD {º FREQUENCY ACCURACY TEST º} 1.006 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.007 HEAD {}ÍÍ FREQUENCY ACCURACY TEST ÍÍ 1.008 DISP Connect the 3325A output to the UUT 1Mê input through 1.008 DISP a 50 ê feedthrough. 1.009 STD Mieco 1402 1.010 JMP 2.001 1.011 EVAL 2.001 IEEE [@3325] FU 1; AM 1 VR; FR 10 MH 2.002 MEMI Use keyboard to enter actual UUT reading in MHz. 2.003 ACC 10.000000MH 0.0000009U 2.004 MEME 2.005 MEMC MH 0.000019U 1V 3.001 IEEE [@3325] AM .001 VR 3.002 HEAD {} 3.003 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 3.004 HEAD {º SELF-CHECK TEST º} 3.005 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 3.006 HEAD {}ÍÍ SELF-CHECK TEST ÍÍ 3.007 DISP Set the UUT front panel controls as follows: 3.007 DISP RANGE switch ................... CHK 3.008 JMP 4.001 3.009 EVAL 4.001 MEMI Use keyboard to enter UUT reading in MHz. 4.002 MATH MEM1=10 4.003 MEME 4.004 MEMC MH 0.000001U 5.001 EVAL {Is the DIR annunciator lite?} 6.001 EVAL {Is the GATE light flashing?} 7.001 HEAD {} 7.002 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍ»} 7.003 HEAD {º RESET TEST º} 7.004 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍͼ} 7.005 HEAD {}ÍÍ RESET TEST ÍÍ 7.006 DISP Push the RESET switch. 7.007 JMP 8.001 7.008 EVAL 8.001 EVAL {Did the display initially show all zeros, then 10MHz?} 9.001 HEAD {} 9.002 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 9.003 HEAD {º RESOLUTION TEST º} 9.004 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 9.005 HEAD {}ÍÍ RESOLUTION TEST ÍÍ 9.006 DISP Slide RESOLUTION switch through all positions to 1MHz, 9.006 DISP noting that one fewer digit is displayed for each step 9.006 DISP of the switch. 9.007 JMP 10.001 9.008 EVAL 10.001 EVAL {Did UUT show proper digits for each switch setting?} 11.001 MEMI Use keyboard to enter UUT reading in GHz. 11.002 MATH MEM1=.01000 11.003 MEME 11.004 MEMC GH 0.0015U 12.001 HEAD {} 12.002 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 12.003 HEAD {º SAMPLE RATE TEST º} 12.004 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 12.005 HEAD {}ÍÍ SAMPLE RATE TEST ÍÍ 12.006 DISP Set RESOLUTION switch to 10Hz. Rotate SAMPLE RATE 12.006 DISP control clockwise and check that the GATE lamp off time 12.006 DISP increases. 12.007 DISP Set SAMPLE RATE control to fully cw position. 12.008 JMP 13.001 12.009 EVAL 13.001 EVAL {Is the GATE lamp off time at least 5 seconds?} 14.001 DISP Set SAMPLE RATE control to HOLD. 14.002 EVAL {Did the GATE lamp stop flashing and is the display held?} 15.001 END